<?xml version="1.0" encoding="iso-8859-1" standalone="no"?>
<!DOCTYPE GmsArticle SYSTEM "http://www.egms.de/dtd/2.0.34/GmsArticle.dtd">
<GmsArticle xmlns:xlink="http://www.w3.org/1999/xlink">
  <MetaData>
    <Identifier>25dga107</Identifier>
    <IdentifierDoi>10.3205/25dga107</IdentifierDoi>
    <IdentifierUrn>urn:nbn:de:0183-25dga1079</IdentifierUrn>
    <ArticleType>Meeting Abstract</ArticleType>
    <TitleGroup>
      <Title language="de">Messtechnischer Vergleich von VSB Audioprozessoren zur intraoperativen Ankopplungskontrolle</Title>
    </TitleGroup>
    <CreatorList>
      <Creator>
        <PersonNames>
          <Lastname>Dziemba</Lastname>
          <LastnameHeading>Dziemba</LastnameHeading>
          <Firstname>Oliver C.</Firstname>
          <Initials>OC</Initials>
        </PersonNames>
        <Address>
          <Affiliation>Universit&#228;tsmedizin Greifswald, Klinik und Poliklinik f&#252;r Hals-, Nasen-, Ohrenkrankheiten, Kopf- und Halschirurgie, Greifswald, Deutschland</Affiliation>
        </Address>
        <Creatorrole corresponding="no" presenting="yes">author</Creatorrole>
      </Creator>
    </CreatorList>
    <PublisherList>
      <Publisher>
        <Corporation>
          <Corporatename>German Medical Science GMS Publishing House</Corporatename>
        </Corporation>
        <Address>D&#252;sseldorf</Address>
      </Publisher>
    </PublisherList>
    <SubjectGroup>
      <SubjectheadingDDB>610</SubjectheadingDDB>
    </SubjectGroup>
    <DatePublishedList>
      <DatePublished>20250318</DatePublished>
    </DatePublishedList>
    <Language>germ</Language>
    <License license-type="open-access" xlink:href="http://creativecommons.org/licenses/by/4.0/">
      <AltText language="en">This is an Open Access article distributed under the terms of the Creative Commons Attribution 4.0 License.</AltText>
      <AltText language="de">Dieser Artikel ist ein Open-Access-Artikel und steht unter den Lizenzbedingungen der Creative Commons Attribution 4.0 License (Namensnennung).</AltText>
    </License>
    <SourceGroup>
      <Meeting>
        <MeetingId>M0607</MeetingId>
        <MeetingSequence>107</MeetingSequence>
        <MeetingCorporation>Deutsche Gesellschaft f&#252;r Audiologie e. V. und ADANO</MeetingCorporation>
        <MeetingName>27. Jahrestagung der Deutschen Gesellschaft f&#252;r Audiologie und Arbeitstagung der Arbeitsgemeinschaft Deutschsprachiger Audiologen, Neurootologen und Otologen</MeetingName>
        <MeetingTitle></MeetingTitle>
        <MeetingSession>Freie Vortr&#228;ge 15: Mittelohr&#47;implantierbare H&#246;rsysteme II</MeetingSession>
        <MeetingCity>G&#246;ttingen</MeetingCity>
        <MeetingDate>
          <DateFrom>20250319</DateFrom>
          <DateTo>20250321</DateTo>
        </MeetingDate>
      </Meeting>
    </SourceGroup>
    <ArticleNo>107</ArticleNo>
  </MetaData>
  <OrigData>
    <TextBlock name="Text" linked="yes">
      <MainHeadline>Text</MainHeadline><Pgraph><Mark1>Hintergrund:</Mark1> Die Qualit&#228;t der Ankopplung des Floating Mass Transducers (FMT) des Mittelohrimplantatsystems Vibrant Soundbrigde (VSB) (MED-EL, &#214;sterreich) ist entscheidend f&#252;r den Versorgungserfolg mit diesem System <TextLink reference="1"></TextLink>. F&#252;r die intraoperative objektive Ankopplungskontrolle sind mehrere Verfahren auf der Basis von Messungen akustisch evozierter Potentiale (AEP) vorgestellt und diskutiert worden <TextLink reference="2"></TextLink>, <TextLink reference="3"></TextLink>. Bei dem Verfahren von Fr&#246;hlich et al. <TextLink reference="4"></TextLink> erfolgte die Wiedergabe des akustischen Stimulus &#252;ber eine modifizierten Audioprozessor des Typs AP404. Dieser Audioprozessortyp ist am Markt nicht mehr verf&#252;gbar.</Pgraph><Pgraph><Mark1>Fragestellung und Methoden:</Mark1> Die Fragestellung dieser Arbeit besteht im messtechnischen Vergleich des AP404 mit aktuell verf&#252;gbaren Prozessoren SAMBA 2 Hi und SAMBA 2 Lo. Hierf&#252;r wurden alle Audioprozessoren mit dem gleichen angenommenen mittleren H&#246;rverlust programmiert. Bei systematischer Beschallung der Ger&#228;te wurde der Ausgangsschallpegel in einer H&#246;rger&#228;temessbox gemessen. Erg&#228;nzend wurde ein weiteres propriet&#228;res Messsystem zur intraoperativen Ankopplungskontrolle in die Vergleichsmessungen mit einbezogen.</Pgraph><Pgraph><Mark1>Ergebnisse:</Mark1> Die Messungen zeigen teils deutliche Unterschiede aller getesteten Prozessoren hinsichtlich Ausgangspegel, Frequenzgang und Dynamik. Ebenso zeigen die Systeme deutlich Unterschiede im Eigenrauschen.</Pgraph><Pgraph><Mark1>Schlussfolgerungen:</Mark1> Die Unterschiede hinsichtlich der Signalverarbeitung der aktuellen VSB-Audioprozessoren k&#246;nnen eine erneute Evaluationsstudie (vergl. Fr&#246;hlich et al. <TextLink reference="4"></TextLink>) notwendig machen. Im Vortrag werden die Messergebnisse detailliert dargestellt und analysiert. Die Eignung des aktuellen propriet&#228;ren Messsystems zur objektiven Ankopplungskontrolle wird diskutiert.</Pgraph></TextBlock>
    <References linked="yes">
      <Reference refNo="1">
        <RefAuthor>M&#252;ller A</RefAuthor>
        <RefAuthor>Mir-Salim P</RefAuthor>
        <RefAuthor>Zellhuber N</RefAuthor>
        <RefAuthor>Helbig R</RefAuthor>
        <RefAuthor>Bloching M</RefAuthor>
        <RefAuthor>Schmidt T</RefAuthor>
        <RefAuthor>Koscielny S</RefAuthor>
        <RefAuthor>Dziemba OC</RefAuthor>
        <RefAuthor>Plontke SK</RefAuthor>
        <RefAuthor>Rahne T</RefAuthor>
        <RefTitle>Influence of Floating-Mass Transducer Coupling Efficiency for Active Middle-Ear Implants on Speech Recognition</RefTitle>
        <RefYear>2017</RefYear>
        <RefJournal>Otol Neurotol</RefJournal>
        <RefPage>809-14</RefPage>
        <RefTotal>M&#252;ller A, Mir-Salim P, Zellhuber N, Helbig R, Bloching M, Schmidt T, Koscielny S, Dziemba OC, Plontke SK, Rahne T. Influence of Floating-Mass Transducer Coupling Efficiency for Active Middle-Ear Implants on Speech Recognition. Otol Neurotol. 2017 Jul;38(6):809-14. DOI: 10.1097&#47;MAO.0000000000001412</RefTotal>
        <RefLink>http:&#47;&#47;dx.doi.org&#47;10.1097&#47;MAO.0000000000001412</RefLink>
      </Reference>
      <Reference refNo="2">
        <RefAuthor>Geiger U</RefAuthor>
        <RefAuthor>Radeloff A</RefAuthor>
        <RefAuthor>Hagen R</RefAuthor>
        <RefAuthor>Cebulla M</RefAuthor>
        <RefTitle>Intraoperative Estimation of the Coupling Efficiency and Clinical Outcomes of the Vibrant Soundbridge Active Middle Ear Implant Using Auditory Brainstem Response Measurements</RefTitle>
        <RefYear>2019</RefYear>
        <RefJournal>Am J Audiol</RefJournal>
        <RefPage>553-9</RefPage>
        <RefTotal>Geiger U, Radeloff A, Hagen R, Cebulla M. Intraoperative Estimation of the Coupling Efficiency and Clinical Outcomes of the Vibrant Soundbridge Active Middle Ear Implant Using Auditory Brainstem Response Measurements. Am J Audiol. 2019 Sep 13;28(3):553-9. DOI: 10.1044&#47;2019&#95;AJA-18-0066</RefTotal>
        <RefLink>http:&#47;&#47;dx.doi.org&#47;10.1044&#47;2019&#95;AJA-18-0066</RefLink>
      </Reference>
      <Reference refNo="3">
        <RefAuthor>Fr&#246;hlich L</RefAuthor>
        <RefAuthor>Rahne T</RefAuthor>
        <RefAuthor>Plontke SK</RefAuthor>
        <RefAuthor>Oberhoffner T</RefAuthor>
        <RefAuthor>Dziemba O</RefAuthor>
        <RefAuthor>Gadyuchko M</RefAuthor>
        <RefAuthor>Hoth S</RefAuthor>
        <RefAuthor>Mir-Salim P</RefAuthor>
        <RefAuthor>M&#252;ller A</RefAuthor>
        <RefTitle>Intraoperative Recording of Auditory Brainstem Responses for Monitoring of Floating Mass Transducer Coupling Efficacy During Revision Surgery-Proof of Concept</RefTitle>
        <RefYear>2020</RefYear>
        <RefJournal>Otol Neurotol</RefJournal>
        <RefPage>e168-e171</RefPage>
        <RefTotal>Fr&#246;hlich L, Rahne T, Plontke SK, Oberhoffner T, Dziemba O, Gadyuchko M, Hoth S, Mir-Salim P, M&#252;ller A. Intraoperative Recording of Auditory Brainstem Responses for Monitoring of Floating Mass Transducer Coupling Efficacy During Revision Surgery-Proof of Concept. Otol Neurotol. 2020 Feb;41(2):e168-e171. DOI: 10.1097&#47;MAO.0000000000002511</RefTotal>
        <RefLink>http:&#47;&#47;dx.doi.org&#47;10.1097&#47;MAO.0000000000002511</RefLink>
      </Reference>
      <Reference refNo="4">
        <RefAuthor>Fr&#246;hlich L</RefAuthor>
        <RefAuthor>Rahne T</RefAuthor>
        <RefAuthor>Plontke SK</RefAuthor>
        <RefAuthor>Oberhoffner T</RefAuthor>
        <RefAuthor>Dahl R</RefAuthor>
        <RefAuthor>Mlynski R</RefAuthor>
        <RefAuthor>Dziemba O</RefAuthor>
        <RefAuthor>Aristeidou A</RefAuthor>
        <RefAuthor>Gadyuchko M</RefAuthor>
        <RefAuthor>Koscielny S</RefAuthor>
        <RefAuthor>Hoth S</RefAuthor>
        <RefAuthor>Kropp MH</RefAuthor>
        <RefAuthor>Mir-Salim P</RefAuthor>
        <RefAuthor>M&#252;ller A</RefAuthor>
        <RefTitle>Intraoperative quantification of floating mass transducer coupling quality in active middle ear implants: a multicenter study</RefTitle>
        <RefYear>2021</RefYear>
        <RefJournal>Eur Arch Otorhinolaryngol</RefJournal>
        <RefPage>2277-2288</RefPage>
        <RefTotal>Fr&#246;hlich L, Rahne T, Plontke SK, Oberhoffner T, Dahl R, Mlynski R, Dziemba O, Aristeidou A, Gadyuchko M, Koscielny S, Hoth S, Kropp MH, Mir-Salim P, M&#252;ller A. Intraoperative quantification of floating mass transducer coupling quality in active middle ear implants: a multicenter study. Eur Arch Otorhinolaryngol. 2021 Jul;278(7):2277-2288. DOI: 10.1007&#47;s00405-020-06313-z</RefTotal>
        <RefLink>https:&#47;&#47;doi.org&#47;10.1007&#47;s00405-020-06313-z</RefLink>
      </Reference>
    </References>
    <Media>
      <Tables>
        <NoOfTables>0</NoOfTables>
      </Tables>
      <Figures>
        <NoOfPictures>0</NoOfPictures>
      </Figures>
      <InlineFigures>
        <NoOfPictures>0</NoOfPictures>
      </InlineFigures>
      <Attachments>
        <NoOfAttachments>0</NoOfAttachments>
      </Attachments>
    </Media>
  </OrigData>
</GmsArticle>